gmi monolith surgical kit consists of an autoclavable case made of high strength resin, containing all the necessary material for the correct positioning of gmi monolith dual connection dental implants system. The components are arranged in such a way to be easy to find in each moment and in any stage of the intervention.
1. Cleaning – Removal of organic material should be performed in an ultrasound basin starting at 24.000 Hz, using enzyme soap (at least 3 enzymes) at the dilution recommended by the manufacturer. Wash with running water and perform a visual check to detect the presence of waste. If positive, repeat the process and should it persist, eliminate by mechanical cleaning. The sterilizations boxes must be cleaned before each use.
2. Conditioning for sterilization – Condition the product on a surgical grade compatible size paper envelope.
3. Moist heat sterilization – shall be conducted in an autoclave without exceeding the maximum values set: 134ºC, 20 minutes, 2.2 bar. Warning – For the sterilization procedure, carefully follow the recommendations in the instruction manual of the autoclave manufacturer.
4. Product life – The useful life of the box will depend on the treatment received during its use.
5. Important notes: