gmi guided surgery kits consist of a sterilizable box by autoclave made of technical plastic, containing all necessary instruments to perform guided surgery procedures for the 4 implant systems frontier, frontier PEAK, avantgard and avantgard PEAK using a new modular tray system:
1. Cleaning – Removal of organic material should be performed in an ultrasound basin starting at 24.000 Hz, using enzyme soap (at least 3 enzymes) at the dilution recommended by the manufacturer. Wash with running water and perform a visual check to detect the presence of waste. If positive, repeat the process and should it persist, eliminate by mechanical cleaning. The sterilizations boxes must be cleaned before each use.
2. Conditioning for sterilization – Condition the product on a surgical grade compatible size paper envelope.
3. Moist heat sterilization – shall be conducted in an autoclave without exceeding the maximum values set: 134ºC, 20 minutes, 2.2 bar. Warning – For the sterilization procedure, carefully follow the recommendations in the instruction manual of the autoclave manufacturer.
4. Product life – The useful life of the box will depend on the treatment received during its use.
5. Important notes:
Presentation of the Guided Surgery kits
Using GMI Frontier GS kit
Using GMI Frontier PEAK GS kit
Using GMI Avantgard GS kit
Using GMI Avantgard PEAK GS kit