Kits

Basic insertion kits

gmi basic insertion kits contain the necessary material for the right insertion of the different implant systems.

Bone compression kit

GMI bone compression kit consists of an autoclavable case made of high strength resin, containing all the necessary to increase implant’s primary stability in type III and type IV low density bones, as well as to improve the treatment at the moment to perform the sinus lifting.

Key features:

  • Case made of high thermal and impact strength technical resin.
  • Security lid locking avoiding accidental opening.
  • Inner supports made of high quality silicon.
  • Extractable inner tray with legs to make easier the manipulation.
  • Bone compressors are machined in high quality stainless steel.Bone compressors have laser depth marks of: 6.5, 8, 10, 11.5 and 13 mm.

Dimensions table

Bone compactors have mechanical marks in the bigger diameter allowing to make an easy selection of the component (1, 2, 3, and 4 marks) and depth laser marks with 6.5, 8, 10, 11.5 and 13 mm implant length (see Figure 1):

Cleaning and Sterilization
1. Cleaning – Removal of organic material should be performed in an ultrasound basin starting at 24.000 Hz, using enzyme soap (at least 3 enzymes) at the dilution recommended by the manufacturer. Wash with running water and perform a visual check to detect the presence of waste. If positive, repeat the process and should it persist, eliminate by mechanical cleaning. The sterilizations boxes must be cleaned before each use.

2. Conditioning for sterilization – Condition the product on a surgical grade compatible size paper envelope.

3. Moist heat sterilization – shall be conducted in an autoclave without exceeding the maximum values set: 134ºC, 20 minutes, 2.2 bar. Warning – For the sterilization procedure, carefully follow the recommendations in the instruction manual of the autoclave manufacturer.

4. Product life – The useful life of the box will depend on the treatment received during its use.

5. Important notes:

  • GMI surgical kits are not supplied in sterile condition and then the user must provide for sterilization before use.
  • Exclusive model for autoclave sterilizing.
  • Clean and sterilize before use.
  • Do not sterilize in an oven as this can cause the box to lose its shape or melt.
  • Avoid the box coming into contact with chemical products as they could cause cracking, staining or general degradation of the product.
  • Carefully follow the guidelines given in the instruction manual of the autoclave manufacturer.

Drills stop kits

GMI drill stop kits consist of an autoclavable technical polymer box with a sliding lid, where the drill stop range for different drill diameters (L1, L2 & L3) are arranged in a simple way to place implants with lenghts between 6.5 and 13 mm. The kits design allow drill stops to be inserted directly on the drill from its housing in the box. Moreover, the kits include the drill stop removal tool as well as a ruler to check the mounted drill stop depth.

Frontier/Avantgard drill kit

Frontier/Avantgard drill kit consists of an autoclavable case made of high strength resin, that allows to complement the PEAK surgical kit by easily extending its use for the placement of 2 additional implant systems.
2 kits for 4 implant systems (Frontier – Avantgard – Frontier PEAK – Avantgard PEAK).

Key features

  • Case made of high thermal and impact strength technical resin.
  • Security lid locking avoiding accidental opening.
    – Inner supports made of high quality silicon.
  • Extractable inner tray with legs to make easier the manipulation.
  • Colour coding for easy identification of drills and drilling protocol for each implant diameter and bone type.
  • Free spare positions for additional instruments.
Kit components

Cleaning and Sterilization
1. Cleaning – Removal of organic material should be performed in an ultrasound basin starting at 24.000 Hz, using enzyme soap (at least 3 enzymes) at the dilution recommended by the manufacturer. Wash with running water and perform a visual check to detect the presence of waste. If positive, repeat the process and should it persist, eliminate by mechanical cleaning. The sterilizations boxes must be cleaned before each use.

2. Conditioning for sterilization – Condition the product on a surgical grade compatible size paper envelope.

3. Moist heat sterilization – shall be conducted in an autoclave without exceeding the maximum values set: 134ºC, 20 minutes, 2.2 bar. Warning – For the sterilization procedure, carefully follow the recommendations in the instruction manual of the autoclave manufacturer.

4. Product life – The useful life of the box will depend on the treatment received during its use.

5. Important notes:

  • GMI surgical kits are not supplied in sterile condition and then the user must provide for sterilization before use.
  • Exclusive model for autoclave sterilizing.
  • Clean and sterilize before use.
  • Do not sterilize in an oven as this can cause the box to lose its shape or melt.
  • Avoid the box coming into contact with chemical products as they could cause cracking, staining or general degradation of the product.
  • Carefully follow the guidelines given in the instruction manual of the autoclave manufacturer.

Frontier/Avantgard GS drill kit

Frontier/Avantgard drill kit consists of an autoclavable case made of high strength resin, containing the necessary drills forcontaining the necessary drills forperforming totally guided surgeries with gmi frontier and avantgard implants of ø3.30 / ø3.75 / ø4.25 and ø4.75 mm and lengths between 8 and 15 mm.

Key features

  • Case made of high thermal and impact strength technical resin.
  • Security lid locking avoiding accidental opening.
  • Inner supports made of high quality silicon.
  • Extractable inner tray with legs to make easier the manipulation.
  • Colour coding for easy identification of drills and drilling protocol for each implant diameter and bone type.
Kit components

Cleaning and Sterilization
1. Cleaning – Removal of organic material should be performed in an ultrasound basin starting at 24.000 Hz, using enzyme soap (at least 3 enzymes) at the dilution recommended by the manufacturer. Wash with running water and perform a visual check to detect the presence of waste. If positive, repeat the process and should it persist, eliminate by mechanical cleaning. The sterilizations boxes must be cleaned before each use.

2. Conditioning for sterilization – Condition the product on a surgical grade compatible size paper envelope.

3. Moist heat sterilization – shall be conducted in an autoclave without exceeding the maximum values set: 134ºC, 20 minutes, 2.2 bar. Warning – For the sterilization procedure, carefully follow the recommendations in the instruction manual of the autoclave manufacturer.

4. Product life – The useful life of the box will depend on the treatment received during its use.

5. Important notes:

  • GMI surgical kits are not supplied in sterile condition and then the user must provide for sterilization before use.
  • Exclusive model for autoclave sterilizing.
  • Clean and sterilize before use.
  • Do not sterilize in an oven as this can cause the box to lose its shape or melt.
  • Avoid the box coming into contact with chemical products as they could cause cracking, staining or general degradation of the product.
  • Carefully follow the guidelines given in the instruction manual of the autoclave manufacturer.

PEAK drill kit

PEAK drill kit consists of an autoclavable case made of high strength resin, that allows to complement the Frontier/Avantgard surgical kit by easily extending its use for the placement of 2 additional implant systems. 2 kits for 4 implant systems (Frontier – Avantgard – Frontier PEAK – Avantgard PEAK).

Key features

  • Case made of high thermal and impact strength technical resin.
  • Security lid locking avoiding accidental opening.
  • Inner supports made of high quality silicon.
  • Extractable inner tray with legs to make easier the manipulation.
  • Colour coding for easy identification of drills and drilling protocol for each implant diameter and bone type.
  • Free spare positions for additional instruments.
Kit components

Cleaning and Sterilization
1. Cleaning – Removal of organic material should be performed in an ultrasound basin starting at 24.000 Hz, using enzyme soap (at least 3 enzymes) at the dilution recommended by the manufacturer. Wash with running water and perform a visual check to detect the presence of waste. If positive, repeat the process and should it persist, eliminate by mechanical cleaning. The sterilizations boxes must be cleaned before each use.

2. Conditioning for sterilization – Condition the product on a surgical grade compatible size paper envelope.

3. Moist heat sterilization – shall be conducted in an autoclave without exceeding the maximum values set: 134ºC, 20 minutes, 2.2 bar. Warning – For the sterilization procedure, carefully follow the recommendations in the instruction manual of the autoclave manufacturer.

4. Product life – The useful life of the box will depend on the treatment received during its use.

5. Important notes:

  • GMI surgical kits are not supplied in sterile condition and then the user must provide for sterilization before use.
  • Exclusive model for autoclave sterilizing.
  • Clean and sterilize before use.
  • Do not sterilize in an oven as this can cause the box to lose its shape or melt.
  • Avoid the box coming into contact with chemical products as they could cause cracking, staining or general degradation of the product.
  • Carefully follow the guidelines given in the instruction manual of the autoclave manufacturer.

PEAK GS drill kit

PEAK drill kit consists of an autoclavable case made of high strength resin, containing the necessary drills for performingcontaining the necessary drills for performingtotally guided surgeries with gmi frontier PEAK and avantgard PEAK implants of ø3.30 / ø3.75 / ø4.25 and ø4.75 mm and lengths between 8 and 15 mm.

Key features

  • Case made of high thermal and impact strength technical resin.
  • Security lid locking avoiding accidental opening.
  • Inner supports made of high quality silicon.
  • Extractable inner tray with legs to make easier the manipulation.
  • Colour coding for easy identification of drills and drilling protocol for each implant diameter and bone type.
Kit components

Cleaning and Sterilization
1. Cleaning – Removal of organic material should be performed in an ultrasound basin starting at 24.000 Hz, using enzyme soap (at least 3 enzymes) at the dilution recommended by the manufacturer. Wash with running water and perform a visual check to detect the presence of waste. If positive, repeat the process and should it persist, eliminate by mechanical cleaning. The sterilizations boxes must be cleaned before each use.

2. Conditioning for sterilization – Condition the product on a surgical grade compatible size paper envelope.

3. Moist heat sterilization – shall be conducted in an autoclave without exceeding the maximum values set: 134ºC, 20 minutes, 2.2 bar. Warning – For the sterilization procedure, carefully follow the recommendations in the instruction manual of the autoclave manufacturer.

4. Product life – The useful life of the box will depend on the treatment received during its use.

5. Important notes:

  • GMI surgical kits are not supplied in sterile condition and then the user must provide for sterilization before use.
  • Exclusive model for autoclave sterilizing.
  • Clean and sterilize before use.
  • Do not sterilize in an oven as this can cause the box to lose its shape or melt.
  • Avoid the box coming into contact with chemical products as they could cause cracking, staining or general degradation of the product.
  • Carefully follow the guidelines given in the instruction manual of the autoclave manufacturer.

Prosthetic kit

gmi prosthetic kit consists of an autoclavable case made of high strength resin, containing all the necessary material for the correct insertion of the whole range of attachments of the gmi implant systems. In addition, the kit includes the abutment extractors for the Avantgard system.

Key features

Case made of high thermal and impact strength technical resin.
Security lid locking avoiding accidental opening.

Inner supports made of high quality colored silicon.

Extractable inner tray with legs to make easier the manipulation.

Clear and natural components distribution.

Color coded silicones according to implant system for they was designed:

  • Blue: Avantgard abutments extraction wrenches.
  • Grey: common instruments.
  • Free socket that allow storing optional instrumental.
Kit components

Cleaning and Sterilization
1. Cleaning – Removal of organic material should be performed in an ultrasound basin starting at 24.000 Hz, using enzyme soap (at least 3 enzymes) at the dilution recommended by the manufacturer. Wash with running water and perform a visual check to detect the presence of waste. If positive, repeat the process and should it persist, eliminate by mechanical cleaning. The sterilizations boxes must be cleaned before each use.

2. Conditioning for sterilization – Condition the product on a surgical grade compatible size paper envelope.

3. Moist heat sterilization – shall be conducted in an autoclave without exceeding the maximum values set: 134ºC, 20 minutes, 2.2 bar. Warning – For the sterilization procedure, carefully follow the recommendations in the instruction manual of the autoclave manufacturer.

4. Product life – The useful life of the box will depend on the treatment received during its use.

5. Important notes:

  • GMI surgical kits are not supplied in sterile condition and then the user must provide for sterilization before use.
  • Exclusive model for
  • Clean and sterilize before use.
  • Do not sterilize in an oven as this can cause the box to lose its shape or melt.
  • Avoid the box coming into contact with chemical products as they could cause cracking, staining or general degradation of the product.
  • Carefully follow the guidelines given in the instruction manual of the autoclave manufacturer.

Short drills kit

gmi short drills kit consists of an autoclavable case made of high strength resin, containing the whole range of drills of the gmi systems with a reduced length, which allow to mill the bone bed in cases with poor mouth opening.

Key features

Case made of high thermal and impact strength technical resin.
Security lid locking avoiding accidental opening.

Inner supports made of high quality colored silicon.

Extractable inner tray with legs to make easier the manipulation.

Clear and natural components distribution.

Color coded silicones according to implant system for they was designed:

  • Blue: Twist drills.
  • Grey: common instruments.
  • Free socket that allow storing optional instrumental.
Kit components

Cleaning and Sterilization
1. Cleaning – Removal of organic material should be performed in an ultrasound basin starting at 24.000 Hz, using enzyme soap (at least 3 enzymes) at the dilution recommended by the manufacturer. Wash with running water and perform a visual check to detect the presence of waste. If positive, repeat the process and should it persist, eliminate by mechanical cleaning. The sterilizations boxes must be cleaned before each use.

2. Conditioning for sterilization – Condition the product on a surgical grade compatible size paper envelope.

3. Moist heat sterilization – shall be conducted in an autoclave without exceeding the maximum values set: 134ºC, 20 minutes, 2.2 bar. Warning – For the sterilization procedure, carefully follow the recommendations in the instruction manual of the autoclave manufacturer.

4. Product life – The useful life of the box will depend on the treatment received during its use.

5. Important notes:

  • GMI surgical kits are not supplied in sterile condition and then the user must provide for sterilization before use.
  • Exclusive model for autoclave sterilizing.
  • Clean and sterilize before use.
  • Do not sterilize in an oven as this can cause the box to lose its shape or melt.
  • Avoid the box coming into contact with chemical products as they could cause cracking, staining or general degradation of the product.
  • Carefully follow the guidelines given in the instruction manual of the autoclave manufacturer.

Thread forms kit

GMI thread form kit consists of an autoclavable case made of high strength resin, containing all the necessary material for the insertion of gmi implant systems in type I and type II high density bones, making easier threading the implant in the bone bed and therefore avoiding thread stress problems.

Key features

  • Case made of high thermal and impact strength technical resin.
  • Security lid locking avoiding accidental opening.
  • Inner supports made of high quality colored silicon.
  • Extractable inner tray with legs to make easier the manipulation.
  • Clear and natural components distribution.
  • Color coded silicones according to implant system for they was designed:
    • Green: thread forms for Phoenix implants.
    • Red: thread forms for Frontier and Avantgard implants.
    • Grey: common thread forms.
  • Thread forms are machined in high quality stainless steel.
  • Thread forms have machined depth marks of: 8, 10, 11.5, 13 and 15 mm.

Kit components

Cleaning and Sterilization
1. Cleaning – Removal of organic material should be performed in an ultrasound basin starting at 24.000 Hz, using enzyme soap (at least 3 enzymes) at the dilution recommended by the manufacturer. Wash with running water and perform a visual check to detect the presence of waste. If positive, repeat the process and should it persist, eliminate by mechanical cleaning. The sterilizations boxes must be cleaned before each use.

2. Conditioning for sterilization – Condition the product on a surgical grade compatible size paper envelope.

3. Moist heat sterilization – shall be conducted in an autoclave without exceeding the maximum values set: 134ºC, 20 minutes, 2.2 bar. Warning – For the sterilization procedure, carefully follow the recommendations in the instruction manual of the autoclave manufacturer.

4. Product life – The useful life of the box will depend on the treatment received during its use.

5. Important notes:

  • GMI kits are not supplied in sterile condition and then the user must provide for sterilization before use.
  • Exclusive model for autoclave sterilizing.
  • Clean and sterilize before use.
  • Do not sterilize in an oven as this can cause the box to lose its shape or melt.
  • Avoid the box coming into contact with chemical products as they could cause cracking, staining or general degradation of the product.
  • Carefully follow the guidelines given in the instruction manual of the autoclave manufacturer.