gmi frontier/avantgard surgical kit (Ref. KYC0F3053) consists of an autoclavable case made of high strength resin, containing all the necessary material for the correct positioning of GMI avantgard coninal connection dental implant system. The components are arranged in such a way as to be easy to find, at all times, the necessary instrument at any time for each stage of the intervention. gmi surgical kits are distributed in cartons, made of high strength technical polymer, and should be steam autoclaved before use at a maximum temperature of 134°C.
Ref. KYC0F3059 - Kit for gmi avantgard system only, without the drill and the wrenches exclusive to the frontier system (see table, highlighted in red).
1. Cleaning – Removal of organic material should be performed in an ultrasound basin starting at 24.000 Hz, using enzyme soap (at least 3 enzymes) at the dilution recommended by the manufacturer. Wash with running water and perform a visual check to detect the presence of waste. If positive, repeat the process and should it persist, eliminate by mechanical cleaning. The sterilizations boxes must be cleaned before each use.
2. Conditioning for sterilization – Condition the product on a surgical grade compatible size paper envelope.
3. Moist heat sterilization – shall be conducted in an autoclave without exceeding the maximum values set: 134ºC, 20 minutes, 2.2 bar. Warning – For the sterilization procedure, carefully follow the recommendations in the instruction manual of the autoclave manufacturer.
4. Product life – The useful life of the box will depend on the treatment received during its use.
5. Important notes: